Secure
- continuous online measurement: measures the dew point of your compressed air via relative humidity and temperature.
- robust stainless steel housing for use in harsh process environments
- precise: high measuring accuracy due to polymer sensor
Sensor technology from BEKO TECHNOLOGIES enables you to accurately measure the degree of humidity or the pressure dew point at critical flow points in your compressed air system. This information is indispensable for operators who want to improve the quality of their compressed air whilst saving energy and costs.
This is particularly important in sensitive areas where very high hygiene standards must be maintained and where compressed air comes into direct or indirect contact with the product. The METPOINT DPM stationary pressure dew point transmitter measures the most critical parameters for compressed air and technical gases i.e. temperature, relative humidity and dew point (down to -60 °Ctd).
But that is not all: The METPOINT DPM sensor provides information regarding the functional capabilities for components and the measurement provides the basic data for the optimum dimensioning for your system. The METPOINT DPM sensor is therefore not just a safety guarantee in your process chain, rather also an instrument for increasing production efficiency.
With the temperature, the relative humidity and the dew point , mobile and stationary dew point measuring devices made by BEKO TECHNOLOGIES precisely measure the critical parameters in compressed air and other gases.
On our YouTube channel you will find more videos about our company, compressed air preparation and measuring technology.
We recommend the delivery as plug & play complete set.
1. Sensor DPM SD23
2. MK-Basic-PN16 measuring chamber with rapid-locking coupling
3. Datenlogger METPOINT BDL data logger compact
Optional: Datenlogger METPOINT BDL data logger portable, Sensor SD11 or SD21
Measuring range | –60 … +30 °Ctd |
Maximum permissible operating overpressure | 50 bar [g] |
Maximum measuring error |
± 4 K für -60 ... -50 °Ctd ± 3 K für -50 ... -30 °Ctd ± 2 K für -30 ... -10 °Ctd ± 1 K für -10 ... +30 °Ctd |
t95 response time |
< 10 sec (from dry to humidity) < 40sec (from humidity to dry) |
Output signal |
4 ... 20 mA, analogue, 4-wire 0 ... 10 V, analogue, 4-wire (special model) RS485, digital output, 4-wire |
Process medium temperature | -30 °C … +70 °C |
Ambient temperature during operation | -25 °C ... +60 °C |
Storage and transport temperature | 40 °C … +85 °C |
Ambient air humidity | 0 … 95 %, no condensation |
MK Basic PN16 | MK Basic PN50 | |
Material | Aluminium | Aluminium |
Maximum permissible operating pressure | 16 bar [g] | 50 bar [g] |
Maximum permissible temperature range | 10 ... 80 °C | 10 ... 80 °C |
* Refer to downloads for additional performance data: Product Folder
Measurement transducers and transmitters are exposed to mechanical stress and temperature fluctuations. As a result e.g. the measuring accuracy of the sensors decreases and generates a so-called “drift”, which then falsifies the measured results and can thereby impair production quality and/or product quality. BEKO TECHNOLOGIES provides a comprehensive range of calibration services for volume flow sensors, pressure dew point transmitters, pressure measurement converters and analyses devices. The calibration is executed according to a specified calibration method from BEKO TECHNOLOGIES and is a so-called ISO calibration. ISO calibration is based on reference measurements performed with externally provided test specimens and certified reference devices. All reference devices can be traced back to national standards. The relevant procedures and methods meet the requirements of the DIN EN ISO 9001 quality standard.